Model T150Hand-heldFrequencyCalibratorOPERATING INSTRUCTIONS
General DescriptionThe T150 Frequency calibrator is designed tobe a high accuracy, simple to use frequencystandard for calibrating process controlinst
2NOTE: The carrier will not honor any claimsunless all shipping material is saved for theirexamination. After examining and removingcontents, save pac
3Operating Modes1. When operating in the output mode the T150has the capability to drive loads up to 1kOhm with TTL levels. The standard outputamplitu
SpecificationsInput/Output Ranges 0 to 1000 CPM 0 to 1000 Hz 0 to 100.0 kHzAccuracy ±0.01% ± 1 countMax Load Driving 5 mA (1 kOhm load min)Input Sign
5MaintenanceGenerally, with normal usage, the calibratorshould hold its rated specifications for at least12 months. Beyond this, it should be returned
02C1977 Rev B 6/11www.martelcorp.come-mail: [email protected]: (603) 434-1433 800-821-0023 Fax: (603) 434-1653 Martel Electronics3 C
Comments to this Manuals