Martel-electronics MS-420 User Manual

Browse online or download User Manual for Equipment Martel-electronics MS-420. Martel Electronics MS-420 User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 1
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 0
milliamps
(+) outputs (–)
4
batt
8 12 16 20
ovl
simulate
source
off
output
select
MINIATURE CURRENT CALIBRATOR
MINI-SOURCE
milliamps
(+) outputs (–)
4
batt
8 12 16 20
ovl
simulate
source
off
output
select
MINIATURE CURRENT CALIBRATOR
MINI-SOURCE
MINI-SOURCE PORTABLE CALIBRATOR
SPECIFICATIONS:
Output: 5 current steps Battery Life: 50 hours @ 4 mA out
(4, 8, 12, 16, 20 mA)
Temperature Range:
Operating: 0° to 50°C
±0.075 of full scale
Accuracy:
Voltage Compliance:
Storage: –25° to 60°C
Source: Internal 9 volt battery
Output Terminations: Two alligator
Simulate: External supply - up to 30 Vdc clips on 8" leads.
Temperature Stability: 100 ppm/°C Size: 3.8" x 2.4" x 1.0"
Power Supply: 9 volt alkaline battery
Unit furnished with standard belt clip.
TYPICAL APPLICATIONS:
CHART RECORDER
PANEL METER
DATA ACQUISITION
OTHER PROCESS INSTRUMENTS
CURRENT SOURCE MODE
CURRENT SIMULATE MODE (2 Wire Transmitter)
Martel Electronics
P.O. Box 770, 1F Commons Dr., Londonderry, NH 03087
Tel: 1-800-821-0023 1-603-434-8179 E-mail: sales@martelcorp.com
4-20 mA LOOP
4-20 mA LOOP
300 ohm
Max. Load
Resistance
30 VDC
Max.
DC
Power
Supply
Mini-Source
Acting as Current Source
Mini-Source
Acting as 2 Wire Transmitter
+
+
+
+
+
Device Under
Calibration
Device Under
Calibration
“The Ultimate Process Loop Tool”
Page view 0
1

Summary of Contents

Page 1 - TYPICAL APPLICATIONS:

milliamps(+) outputs (–)4batt8 12 16 20ovlsimulatesourceoffoutputselectMINIATURE CURRENT CALIBRATORMINI-SOURCEmilliamps(+) outputs (–)4batt8 12 16 20o

Comments to this Manuals

No comments